1、精选优质文档-倾情为你奉上附录一:英文技术资料翻译英文原文:Emerg Infect Dis. 2008 August; 14(8): 12551258.doi: 10.3201/eid1408.PMCID: PMCCutaneous Infrared Thermometry for Detecting Febrile Patients, , , , and * This article has been other articles in PMC.AbstractWe assessed the accuracy of cutaneous infrared thermometry, which
2、 measures temperature on the forehead, for detecting patients with fever in patients admitted to an emergency department. Although negative predictive value was excellent (0.99), positive predictive value was low (0.10). Therefore, we question mass detection of febrile patients by using this method.
3、Keywords: Fever, mass detection, cutaneous infrared thermometry, infectious diseases, emergency, dispatchRecent efforts to control spread of epidemic infectious diseases have prompted health officials to develop rapid screening processes to detect febrile patients. Such screening may take place at h
4、ospital entry, mainly in the emergency department, or at airports to detect travelers with increased body temperatures (). Infrared thermal imaging devices have been proposed as a noncontact and noninvasive method for detecting fever (). However, few studies have assessed their capacity for accurate
5、 detection of febrile patients in clinical settings. Therefore, we undertook a prospective study in an emergency department to assess diagnostic accuracy of infrared thermal imaging.The StudyThe study was performed in an emergency department of a large academic hospital (1,800 beds) and was reviewed
6、 and approved by our institutional review board (Comit de Protection des Personnes se Prtant la Recherche Biomdicale Piti-Salptrire, Paris, France). Patients admitted to the emergency department were assessed by a trained triage nurse, and several variables were routinely measured, including tympani
7、c temperature by using an infrared tympanic thermometer (Pro 4000; Welch Allyn, Skaneateles Falls, NY, USA), systolic and diastolic arterial blood pressure, and heart rate.Tympanic temperature was measured twice (once in the left ear and once in the right ear). This temperature was used as a referen
8、ce because it is routinely used in our emergency department and is an appropriate estimate of central core temperature (). Cutaneous temperature was measured on the forehead by using an infrared thermometer (Raynger MX; Raytek, Berlin, Germany) (). Rationale for an infrared thermometer device instea
9、d of a larger thermal scanner was that we wanted to test a method (i.e., measurement of forehead cutaneous temperature by using a simple infrared thermometer) and not a specific device. The forehead region was chosen because it is more reliable than the region behind the eyes (,). The latter region
10、may not be appropriate for mass screening because one cannot accurately measure temperature through eyeglasses, which are worn by many persons. Outdoor and indoor temperatures were also recorded.Measurement of cutaneous temperature with an infrared thermometer. A) The device is placed 20 cm from the
11、 forehead. B) As soon as the examiner pulls the trigger, the temperature measured is shown on the display. Used with permission.The main objective of our study was to assess diagnostic accuracy of infrared thermometry for detecting patients with fever, defined as a tympanic temperature 38.0C. The se
12、cond objective was to compare measurements of cutaneous temperature and tympanic temperature, with the latter being used as a reference point. Data are expressed as mean standard deviation (SD) or percentages and their 95% confidence intervals (CIs). Comparison of 2 means was performed by using the
13、Student t test, and comparison of 2 proportions was performed by using the Fisher exact method. Bias, precision (in absolute values and percentages), and number of outliers (defined as a difference 1C) were also recorded. Correlation between 2 variables was assessed by using the least square method.