血糖仪比对记录表.doc
《血糖仪比对记录表.doc》由会员分享,可在线阅读,更多相关《血糖仪比对记录表.doc(8页珍藏版)》请在淘文阁 - 分享文档赚钱的网站上搜索。
1、Four short words sum up what has lifted most successful individuals above the crowd: a little bit more.-author-date血糖仪比对记录表血糖仪比对记录表血糖仪比对记录表时间: 2017 年 9 月8日标准:血糖4.2mmol/L,允许偏差在0.83mmol/L以内;血糖浓度4.2mmol/L,允许偏差20%以内。生化仪型号:东芝FX-8样本1测定值:2.02 mmol/L允许范围:1.19-2.85mmol/L样本2测定值:7.76 mmol/L允许范围:6.21-9.31mmol/L
2、样本3测定值:3.40 mmol/L 允许范围:2.57-4.23mmol/L样本4测定值:18.10 mmol/L允许范围:14.48-21.72mmol/L比对科室血糖仪测定结果分数结论样本1样本2样本3样本4儿科1.907.303.5017.50100合格儿科1.907.403.3017.80100合格手术室1.406.702.9016.10100合格神经内一科1.506.803.0016.90100合格产二科2.307.404.1019.70100合格五官科2.407.203.8017.80100合格妇科2.407.703.8019.30100合格呼吸消化科2.207.303.7016
3、.90100合格呼吸消化科2.207.203.5018.20100合格综合内科2.307.703.8018.20100合格综合内科2.207.603.6018.20100合格神经内二科2.307.703.8019.20100合格神经内二科2.307.903.8018.60100合格心内科2.106.903.5017.80100合格产一科2.107.603.7017.60100合格急诊科2.107.403.4017.40100合格急诊科2.107.103.3017.70100合格急诊科2.007.203.3016.60100合格急诊科2.107.303.3016.60100合格骨科1.907.203.3017.80100合格对比结论: 本次对比普通外科、ICU因故未参加,其余参加科室均为合格。 我院目前全部使用京都品牌的同一款血糖仪,方面管理和结果比较。-
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- 血糖仪 记录
限制150内