1、Four short words sum up what has lifted most successful individuals above the crowd: a little bit more.-author-date血糖仪比对记录表(1)血糖仪比对记录表血糖仪比对记录表时间: 2017 年 9 月8日标准:血糖4.2mmol/L,允许偏差在0.83mmol/L以内;血糖浓度4.2mmol/L,允许偏差20%以内。生化仪型号:东芝FX-8样本1测定值:2.02 mmol/L允许范围:1.19-2.85mmol/L样本2测定值:7.76 mmol/L允许范围:6.21-9.31mmo
2、l/L样本3测定值:3.40 mmol/L 允许范围:2.57-4.23mmol/L样本4测定值:18.10 mmol/L允许范围:14.48-21.72mmol/L比对科室血糖仪测定结果分数结论样本1样本2样本3样本4儿科1.907.303.5017.50100合格儿科1.907.403.3017.80100合格手术室1.406.702.9016.10100合格神经内一科1.506.803.0016.90100合格产二科2.307.404.1019.70100合格五官科2.407.203.8017.80100合格妇科2.407.703.8019.30100合格呼吸消化科2.207.303.7
3、016.90100合格呼吸消化科2.207.203.5018.20100合格综合内科2.307.703.8018.20100合格综合内科2.207.603.6018.20100合格神经内二科2.307.703.8019.20100合格神经内二科2.307.903.8018.60100合格心内科2.106.903.5017.80100合格产一科2.107.603.7017.60100合格急诊科2.107.403.4017.40100合格急诊科2.107.103.3017.70100合格急诊科2.007.203.3016.60100合格急诊科2.107.303.3016.60100合格骨科1.907.203.3017.80100合格对比结论: 本次对比普通外科、ICU因故未参加,其余参加科室均为合格。 我院目前全部使用京都品牌的同一款血糖仪,方面管理和结果比较。-