PiCCO在ICU中的应用.ppt
《PiCCO在ICU中的应用.ppt》由会员分享,可在线阅读,更多相关《PiCCO在ICU中的应用.ppt(61页珍藏版)》请在淘文阁 - 分享文档赚钱的网站上搜索。
1、PiCCO容量监护仪在ICU中的应用,2,PiCCO Technology,容量管理,Introduction to the PiCCO-Technology,CO,Volumetric preload,EVLW,Contractility,Differentiated Volume Management,static - dynamic,3,正确的监测才能进行正确的治疗,4,什么是PiCCO?,5,PiCCO是一种技术,是一种简便、微创、高效费比的,对重症病人主要血流动力学参数进行监测的工具。,6,Central venous catheter jugular subclavian femo
2、ral,大多数血流动力学不稳定的患者都会置 中心静脉导管 (vasoactive agents administration) 动脉导管 (monitoring of blood pressure, blood samples),Thermodilution arterial catheter femoral axillary long radial,连接,7,Central venous catheter jugular subclavian femoral,8,PiCCO技术可以监测心输出量,9,PiCCO Technology is a combination of transpulmo
3、nary thermodilution and pulse contour analysis,Principles of Measurement,Left Heart,Right Heart,Pulmonary Circulation,Lungs,Body Circulation,PULSIOCATH,PULSIOCATH,CVC,PULSIOCATH arterial thermodilution catheter,central venous bolus injection,Introduction to the PiCCO-Technology Function,10,Transpulm
4、onary thermodilution method,Arterial TD catheter 动脉热稀释导管 (e.g. PV2014L16),CV bolus injection 中心静脉一次注射,RAEDV 右房舒张末容量,PBV 肺血容量,EVLW,LAEDV 左房舒张末容量,LVEDV 左室舒张末容量,EVLW 血管外肺水,RVEDV 右室舒张末容量,0,1,0,2,0,3,0,4,0,5,0,0,0,0,2,0,4,0,6,C,-,D,T,Injection,s,Lungs,CVC,Aorta,V. cavasup.,RA,Arterial TD catheter,11,Tb =
5、 Blood temperature Ti = Injectate temperature Vi = Injectate volume Tb . dt = Area under the thermodilution curve K = Correction constant, made up of specific weight and specific heat of blood and injectate,心输出量是利用 Stewart-Hamilton 公式对热稀释曲线进行分析所得,热稀释方法所得心输出量的计算,12,The area under the thermodilution c
6、urve is inversely proportional to the CO.,36,5,37,5,10,Thermodilution curves,Normal CO: 5.5l/min,Introduction to the PiCCO-Technology Thermodilution,36,5,37,36,5,37,Time,low CO: 1.9l/min,High CO: 19l/min,Time,Time,Temperature,Temperature,Temperature,13,经肺 vs. 肺动脉热稀释,Left heart,Right Heart,Pulmonary
7、Circulation,Lungs,Body Circulation,PULSIOCATH arterial thermo-dilution catheter,central venous bolus injection,RA,RV,PA,LA,LV,Aorta,Transpulmonary TD (PiCCO),Pulmonary Artery TD (PAC),这两种方法所得CO都是准确的,因为热稀释方法关注的只是 时间以及温度改变!,14,Transpulmonary thermodilution CO measurement (vs PA thermodilution),Author
8、Bias (L/min) SD r Godje Chest 1998 0.16 0.35 0.96 Sakka ICM 1999 0.68 0.62 0.97 Goedje CCM 1999 0.29 0.60 0.93 Bindels CC 2000 0.49 0.78 0.95 Goedje Chest 2000 0.35 0.72 0.98 Della Rocca BJA 2002 0.15 0.87 0.93 Sander CC 20050.000.70 0.95 Ostergaard AAS 2006 0.46 0.55,15,Pulse contour analysis 脉搏轮廓分
9、析,Cal = calibration factor obtained from transpulmonary thermodilution (cold bolus injection) 校准系数来自经肺热稀释方法,Surface = Cal. x Stroke volume,P (mmHg),t (s),17,Validation of continuous cardiac output measurement by the pulse contour analysis (PiCCO system) PiCCO所得连续心排量数值准确性论证,Roedig G et al. Br J Anaes
10、th 1999; 82: 525-530 Goedje O et al. Ann Thorac Surg 1999; 68: 1532-1536 Buhre W et al. J Cardiothorac Vasc Anesth 1999; 13: 437-440 Goedje O et al. Crit Care Med 1999; 27: 2407-2412 Zollner C et al. J Cardiothorac Vasc Anesth 2000;14: 125-129 Goedje O et al. Med Sci Monit 2001; 7: 1344-1350 Felbing
11、er TW et al. J Clin Anaesth 2002; 14: 296-301 Goedje O et al. Crit Care Med 2002; 30:52-58 Rauch H et al. Acta Anaesthesiol Scand 2002; 46: 426-429 Felbinger et al. J Clin Anaesth 2005; 17: 241-248 Ostergaard et al. Acta Anaesthesiol Scand 2006; 50: 1044-1049,18,PiCCO不仅是 心排量监测仪,19,PiCCO technology,G
12、EDV评估心脏前负荷容量,20,目前医学界判斷心脏前負荷的标准,Centrol venous pressure (CVP) Pulmonary capillary wedge pressure (PCWP) 但是 這是基於以下的假設 “ 压力 = 容量 但這種假設並非完全正確,PULSION Medical Systems,21,Kumar et al., Crit Care Med 2004;32: 691-699,21,灌注压 CVP / PCWP反映前负荷,中心静脉压和每搏输出量的关联,监测前负荷,22,Kumar et al., Crit Care Med 2004;32: 691-6
13、99,22,肺动脉嵌压和每搏输出量的关联,Measuring Preload,监测前负荷,灌注压 CVP / PCWP反映前负荷,23,Raper R and Sibbald WJ. Chest 1986 Misled by the wedge ? The Swan-Ganz catheter and left ventricular preload.,24,GEDV可以作为心脏前负荷的指标,Global End-Diastolic Volume evaluated by transpulmonary thermodilution,GEDV,25,CO x MTt = volume of di
14、stribution = 1 + 2 + 3 + 4 + 5,3,Indicator injection,Indicator dilution curve,26,CO x MTt = volume of distribution = 1 + 2 + 3 + 4 + 5,3,CO x Dst = largest mixing chamber = 3,Indicator injection,Indicator dilution curve,27,CO x MTt = volume of distribution = GEDV + PBV + EVLW,EVLW,Meier et al. J App
15、l Physiol 1954,28,Changes in stroke volume index (%),Changes in global end-diastolic volume index (%),r = 0.72 p 0.001,Transpulmonary thermodilution as a guide to fluid therapy in septic patients with acute circulatory failure. Michard et al. SCCM 2002,29,治疗指南Surviving Sepsis Campaign Guidelines 200
16、8: 灌注压在评估扩容治疗中已经被证实有局限性 . 使用血流量或者容量参数指导容量管理或许更有优势 这些技术已经能够在ICU内实现 ,PiCCO技术,30,Intrathoracic pressure Venous return to left and right ventricle Left ventricular preload Left ventricular stroke volume Systolic arterial blood pressure,Intrathoracic pressure Squeezing “ of the pulmonary blood Left ventr