《dsc测量比热容》PPT课件.pptx
《《dsc测量比热容》PPT课件.pptx》由会员分享,可在线阅读,更多相关《《dsc测量比热容》PPT课件.pptx(20页珍藏版)》请在淘文阁 - 分享文档赚钱的网站上搜索。
1、New proposed methodology for specific heat capacity determination of materials for thermal energy storage(TES)by DSCDSC Abstract5.Conlusions ABSTRACTThis study presents a methodology to determine the specific heat the specific heat capacitycapacity(Cp)of materials for thermal energy storage(TES)by D
2、SC.Three different DSC methods are considered to be applied in the methodology,and are explained and compared in this study in order to select the most proper one for Cp determination.To perform this study,the Cp of three materials commonly used in sensible TES systems,slate,water,and potassium nitr
3、ate(KNO3),is determined.一、INTRODUCTIONllack of fossil energetic resourceslits pricelneed to decrease CO2 emissionslead to dedevelop new energy systems able to take profit of renewable energy and be environmentally friendly and less expensiveTES systems are a possible solutionSolar energy abundant cl
4、ean and easy accessible.problem:as the hours of maximum energy demand match with the hours of no solar irradiationwith the aim to fill this existing energy supply gap,TES systems are presented as the solution to store the energy and use it anytimeTES technologieslsensible heat storage the process by
5、 which the heat is accumulated due to the increase of the material temperature without experimenting structural changes,no phase changellatent heat storagephase change,thermochemical energymaterial selectionA high storage capacity Therefore,need a high specific heat capacity.(1)定义:在定义:在程序控制温度程序控制温度下
6、,测量下,测量物物质与参比物之间的能量差质与参比物之间的能量差随随温度变温度变化关系化关系的一种技术。的一种技术。DSC曲线峰包围的面积正比于热焓曲线峰包围的面积正比于热焓的变化的变化可以测定多种热力学和动力学参数,可以测定多种热力学和动力学参数,例如比热容、反应热、转变热、相例如比热容、反应热、转变热、相图、反应速率、结晶速率、高聚物图、反应速率、结晶速率、高聚物结晶度、样品纯度等。该法使用温结晶度、样品纯度等。该法使用温度范围宽(度范围宽(-175725)、分辨率)、分辨率高、试样用量少。适用于无机物、高、试样用量少。适用于无机物、有机化合物及药物分析。有机化合物及药物分析。差示扫描量热法
7、差示扫描量热法(DSC)(Differential Scanning Calorimetry)采用试样和一标准物质在相同条件下进行扫描,然后量出二者的纵坐标进行计算。要求标准物质在所测温度范围内没有发生物理化学变化,且比热容已知。常用的标准物质是蓝宝石。具体作法是在DSC内先用两个空样品皿,以一定的升温速度作一条基线,然后放入蓝宝石样品在相同条件下作一条DSC曲线,最后在同样条件下作未知试样的DSC曲线,如图three DSC methods for Cp measurementlThe dynamic method has traditionally been used to measure
8、 PCM latent heat but also for Cp measurements applying high heating rates in the temperature range of study lThe areas method is based on the Cp value,thus,how much heat flux is needed to heat up a material for temperature increase,and consists of consecutive isothermal stages differing 1 with no he
9、ating segments amidlThe isostep method is an intermediate between these two,as it is a succession of dynamic methods applied to heat up the material just 1 between isothermal segments.This method has been tested in glass transition pharmaceutical studies with good results obtained,but it has not bee
10、n used for TES materials Cp determination,thereby its selection for this study二、二、MaterialsTo ensure the performance of the methodology in a representative variety of material phase forms and chemical structures,three different materials widely used in sensible heat storage systems have been chosen
11、to perform this study.lWater.Its thermal properties are well known,and therefore,these values can be used as a reference to see the approach on the measurements each method has.Commercial Bi-distillated water from Panreac has been used to perform the experimentslSlate.It is a widely used constructio
12、n material and its performance depends on its specific heat capacity,therefore authors find it an interesting material to be tested.The slate samples used in this study were taken from a quarry in the Catalan Pyrenees.lPotassium nitrate.It is an inorganic salt mainly used as molten salt in concentra
13、ted solar power plants(CSP Plants).It has high melting point(320)and,as all inorganic salts,a complex chemical structure,and that is why it is also found to be useful for this papers purpose.It is important to remark that this material selection responds to the need to test the methodology and not t
14、o tie it to a concrete material type or a specific TES application.Therefore,materials used in different TES systems and with which authors commonly work have been selected.determined by differential scanning calorimetry(DSC)Standard 40 uL aluminium crucibles weigh with a analytical balance with a p
15、recision of 0.01 mg.The procedure to determine the specific heat capacity of a material consists of three different measurements,all done under the same conditions,thus,using the same DSC method:Blank measurement.It is necessary to run an experiment with an empty crucible to measure the heat flux th