1、13 .峰面积波动Q.:What are the reasons for fluctuations in peakarea, when the same sample is injected multipletimes?问:同一样品多次注入时,峰面积波动的缘由是什么?A.:Unfortunately, there are many different reasons, too many to count. Nearly every part of the HPLC instrument can conceivably contribute to changes in peak area. Le
2、t us review them one by one.答:不幸的是,有很多不同的缘由。HPLC仪器的每个部分都可以导致峰面积的变化。让我 们一一检查。The inject or comes to mind first. What problems you might encounter depends to some degree on the type of injector. But in all cases, the formation of air bubbles during the measurement of the injection volume is a major pr
3、oblem source.If you areinjecting manually,makesure that thereare no air bubbles in your syringe. With fixed-loop injectors,make sure that no air is siphoned into theinjector loop. Inautoinjectors, airbubbles can be formed if the cap of the vial seals well around the needle,which can give rise to a v
4、acuum when a large portion of the sample is injected. Any air bubble form ation results in a random variation of the injection volume from injection to injection.首先想到进样器。您可能遇到的问题在某种程度上取决于进样器的类型。但是在全部状况下, 在测量进样体积期间气泡的形成是主要的问题来源。假如您是手动进样,请确保注射器中没 有气泡。使用固定回路进样器时,确保没有空气进入进样器环路。在自动进样器中,假如样 品瓶的盖子密封在针头四周则会
5、形成气泡,当注入大部分样品时会产生真空。任何气泡的形 成都会导致每次进样之间进样量的差异。sample and the peak areaobtainedfrom a singlesample becomes constantonly after two or three injections,samplecarry-overis a likely source of the problem. Cleanthe syringe carefully after each injection, if you are injectingmanually, or check the needle wa
6、sh in anauto injector. Makesure that thesolvent used for the needle wash is a goodsolventfortheanalyte(s).假如样品之间分析物的浓度差异很大,并且从单个样品获得的峰面积仅在两次或三次进样后 才保持恒定,样品残留可能是问题的根源。每次注射后(假如您是手动注射)请认真清洗注 射器,或在自动进样器中检查针头清洗。确保用于针头清洗的溶剂是分析物的良好溶剂。Decreasesin peak area from run to run can also be causedby temperature ch
7、anges, but the effect is small, under 2%. Ifyou take a sample out of the refrigeratorand put it into anautoinjectorjtwillslowlywarmuptotheinjectortemperature and the solvent will expand.Therefore you willinject initially alarger mass of sample than later when thesamplehasreachedthetemperatureofthesa
8、mplecompartment.每次运行时峰面积的削减也可能是由于温度变化引起的,但是影响很小,不至IJ 2%。假如将 样品从冰箱中取出并放入自动进样盘中,它将缓慢升温至进样器温度,并且溶剂会膨胀。因 此,当样品达到样品盘的温度时,您开头时注入的样品体积比随后的大。Randomvariationsoftheflowratecanalsocausefluctuationsinpeakarea.Theyaretypicallycaused by malfunctioning check valves or air orcavitation in the pumphead. They are usu
9、ally accompanied by fluctuations in retention times aswell. You can check if this is the problem by monitoring the backpressure. What fluctuations can betolerated depends onthe width of the peak. Assume thatthereis a flow difference of 10% between the two heads ofyour pump. If your peak width is abo
10、ut 20 pump strokes, the fluctuation in peak area caused by the pulsation of the pumpisunderl%. But it will be 10% if the peak width is aboutone pumpstroke.流速的随机变化也会导致峰面积波动。它们通常是由止回阀故障或泵头中的空气或气蚀引起 的。它们通常还伴随着保留时间的波动。您可以通过监控背压来检查这是否是问题所在。可 以容忍的波动取决于峰的宽度。假设您的泵的两个扬程之间的流量差为10%。假如您的峰 宽约为20个泵冲程,则由泵的脉动引起的峰面积
11、波动小于1%。但是,假如峰值宽度约为 一个泵冲程,则它将为10%。Anotherpotential cause of flow-rate changes are leaks onthe high-pressure side of the system. They should be easy tofind.流速变化的另一个潜在缘由是系统高压侧泄漏。应当很简洁找到。Integrationerrors are more difficult to troubleshoot. Grosschanges in the way the software integrates the peaks can b