1、祛码使用操作规范Document Issue/Revision Status:文件发布/修订描述:Revision版本Release/revise Date 发布/修订日期Revise Outline制修纪要Author 作者Remark 备注Drafted by制订Checked by 审核Approved by 批准Department制订部门1. Purpose 目的Provide the operation and maintenance methods of the weight for the user. By ensure accurate, reliable, stable o
2、f weight, and extend its service life, to ensure the quality of products and measurement accuracy of results.为使用者提供祛码的正确使用和保养方法,确保祛码的量值准确、可靠、稳定和延长其使用寿命,从 而保证产品质量及测量结果的准确度。2. Range 范围For the company 9s in-serviceweights.适用于本公司所有在用型号的祛码。3. Authorities and responsibilities 权责1. Quality Department calib
3、ration center is responsible for the measurement, repair and the calibrationafter repairedof weight.Guide the operator use and maintenance the device properly.3.1. 品质部计量负责祛码的校准、维修和维修后再校准并指导操作者正确使用及保养。3. 2. The user department is responsible for the correct use of weights, normal maintenance and subm
4、ission on schedule.3.2. 各使用部门负责祛码的正确使用、日常检查、维护保养及到期送检。4. content 内容4. L The working environment:4.L 使用环境:Temperature fluctuations less than 1 during working, relative humidity less than 80% RH. 使用期间温度波动小于1,相对湿度小于80%RH。4. 2. Instructions:1. 2.使用方法:4. 2. 1. Put on cotton gloves, steady hold.5. 2.L 戴上棉
5、纱手套,拿取时应稳拿轻放。6. 2. 2. After use, wipe the weight clean, put back into the storage box.7. 2. 2.用完之后将祛码擦拭干净放回存放盒内。8. Inspection and maintenance 点检和保养5. 1. Inspection method。5. L点检方法:5. L L weight should bepasted with a label, no damage to the label, shedding and other phenomena, and ia the period of v
6、alidity.If exceed the validity period , need send (o thn Quality Department to have re-test, calibration, labeling before use.6. LI.祛码应贴有标签,标签无损坏、脱落等现象,并且在有效期内,假设超过有效期需由品质部重新测 试、校准、贴标后才可使用。7. 1. 2. Count the number of weight, checkthe surface whether have visible defects, then wipe clean。8. L2.检查祛码是
7、否齐全,并El测祛码的测量面有无目视可见的缺陷,然后将其擦拭干净。5.2. Maintenance:5.2.保养方法:5 2. L Must wear gloves When using thn weight, cant touch the weight of the work surfacedircctly, to avoid corrosion.5.2.1 .使用祛码时必须戴上细纱手套,不可用手直接接触祛码的工作面,以免锈蚀。5. 2. 2. Weights should avoided collision, scratches, near the absence of vibration
8、, also ths relative movementduring the process of measurement.5.2.2. 硅码在使用过程中应防止碰撞、划伤,附近应没有振动,还应防止测量中的相对运动。5. 2. 3. End of use, clean the weight according to the provisions, and then placed into the balance room. The temperature of each level weight 再Iso grade 1 to grade 4 should be the same as the
9、room temperature. The test-room waitingtime for measurement of level 1 and grade 1 should not less than 24 hours , grade 2 and grade 2, grade 4 and grade 4 shall not less than 12 hours, 5/6 grade weight can be checked after cleaning.5.3.3. 祛码使用完毕后按规定清洗,然后放在天平室内。各等和1级至4级祛码的温度应与室温相一致, 一一等和1级、2级祛码在检定空内
10、存放时间不得少于24小时丁二等和3级、4级祛码的存放时 间不得少于12小时,5/6级大祛码清洗晾干后即可检定。6. Precautions,考前须知1. All levels of weight surfaces should be smooth, clean and tidy, there shall be no obvious trachoma, cracks, burrs ,sharp edges and other defects, The surface of each level and grade 1 to 4weight should be well polished.6.1.
11、 各级祛码外表均应光滑、平整洁净,不得有显见的砂眼、裂纹、毛刺和锐边等缺陷,各等祛码及1 至4级祛码的外表应有良好的抛光。6. 2. weight surface coating or covering allow to protect the weight surface from corrosion. The coating or covering must withstand shock, wear and atmospheric conditions. Have uniform surface color, without delamination, blistering, cracks, bums, flow marks, wrinkles, cracks, exposed defects.6.2. 祛码外表允许有镀层或涂层,以保护祛码外表不受腐蚀。此镀层或涂层必须能承受冲击、磨损及大 气环境的影响,外表色泽均匀,不得有脱层、起泡、裂纹、灼伤、流痕、起皱、龟裂、露底等缺陷。7. Figure 附图
限制150内