1、CT 成像原理介紹CT 成像原理介紹Computed Tomography CT Basics Principle of Spiral CT Scan Parameter&Image Quality Optimizing Injection Protocols Clinical ApplicationsComputed Tomography CT BasicsX-Ray Discovery X-ray was discovered by a German scientist Roentgen 100 years ago.This made people for the first time b
2、e able to view the anatomy structure of human body without operation77 But its superimposed 77 And we couldnt view soft tissueX-Ray Discovery X-ray was disl My name is Godfrey Hounsfieldl I work for the Central Research Labs.of EMI,Ltd in Englandl I developed the the first clinically useful CT scann
3、er in 1971History of Computed Tomography Early 1970s1963-Alan Cormack developed a mathematical method of reconstructing images from x-ray projections My name is Godfrey Hounsfiel For the first time we could view:CT Broke the Barrier-Tomographic or“Slice”anatomy-Density difference77 But its time cons
4、uming77 And resolution needs to be improved For the first time we could vConcept of X-ray AttenuationAn X-ray beam passing throughthe body is attenuated(loses itsenergy)by:AbsorptionScatteringIncident X-rayTransmitted raySCATTERED X-SCATTERED X-RAYSRAYSBODYTISSUEAbsorption by the tissue is proportio
5、nal to the densityLessdensetissueMoredensetissueMORE ATTENUATIONLESS ATTENUATIONConcept of X-ray AttenuationAHow does CT Work?How does CT Work?Recon.&postpro.Data acquisitionX-ray generationHow does CT Work?Recon.&postX-ray goes through collimator thereforepenetrate only an axial layer of the object
6、,called sliceHow does CT Work?How does CT Work?X-ray goes through How does CTPatient is placed in the center of the measurement fieldX-ray is passed through the patients slice from many direction along a 360 pathThe transmitted beams are captured by the detectors which digitizes these signalsThese d
7、igitized signals called raw data are sent to a computer which create the CT imageHow does CT Work?How does CT Work?Patient is placed in the centeThe object slice is divided into small volume elements called voxels.Each voxel is assigned a value which is dependent on the average amount of attenuation
8、How is CT Image generated?The object slice is divided inThe attenuation values are transferred to the computer where they are coded&used to create a slice imageHow is CT Image generated?The attenuation values are traCT Generations&Design“Generation”is used to label CT tube-detector designs3rd Genera
9、tion Design3rd Generation DesignRotating tube&detectorRotating tube&detector4th Generation Design4th Generation DesignFixed ring detectorFixed ring detectorCT Generations&Design“GeneraSlip-ring TechnologyPower is transmitted through parallel sets of conductive ringsinstead of electrical cables Conti
10、nuous Gantry Rotation Prerequisite for Spiral CTNon Slip-ring ScannerNon Slip-ring ScannerSlip-ring ScannerSlip-ring ScannerSlip-ring TechnologyPower is tComputed Tomography CT Basics Principle of Spiral CT Scan Parameter&Image Quality Optimizing Injection Protocols Clinical ApplicationsComputed Tom
11、ography CT BasicsWhat is Spiral Scan?-just 4“C”Continuously rotating tube/detector systemContinuously generating X-rayContinuously table feedContinuously data acquisitionWhat is Spiral Scan?-just 4Reconstruction of arbitraryslices(either contiguous or overlapping)within thescanned volume Distance be
12、tween the slices is called IncrementABVolume DataContinuous data acquisitionReconstruction of arbitraryABVIncrement Slice ThicknessIncrement=Slice ThicknessIncrement=Slice Thickness No Overlap No Overlap No Gaps No GapsContiguous Image ReconstructionIncrementSlice ThicknessIncrIncrement Overlap Slic
13、eThicknessOverlapping Image ReconstructionIncrement Slice ThicknessIncrement Slice ThicknessIncrement Slice Thickness Gaps between slices Gaps between slices Images are further apart Images are further apart Less images created Less images createdImage Reconstruction with GapsIncrement Slice Thickne
14、ssIncrement Slice ThicknessImaStandard CT/Slice ImagingStandard CT/Slice ImagingDeep InspirationDeep InspirationShallow InspirationShallow Inspiration Misregistration due to different Misregistration due to different respiratory levels between slices respiratory levels between slicesPartial Volume E
15、ffectPartial Volume Effect Unable to resconstruct images at Unable to resconstruct images at arbitrary position arbitrary position Slice imaging is slow Slice imaging is slowStandard CT/Slice ImagingDee Scan the whole region of Scan the whole region of interest in one breath hold interest in one bre
16、ath hold Reconstruction of overlapping Reconstruction of overlapping images without additional dose images without additional dose Retrospective reconstruction Retrospective reconstruction of slices in arbitrary position of slices in arbitrary position within the scanned volume within the scanned vo
17、lume No gaps since radiation always No gaps since radiation always transmits the whole volume transmits the whole volumeSpiral CT/Volume ImagingSpiral CT/Volume Imaging Scan the whole region of ReComputed Tomography CT Basics Principle of Spiral CT Scan Parameter&Image Quality Optimizing Injection P
18、rotocols Clinical ApplicationsComputed Tomography CT BasicsScan ParametersX-ray Tube Voltage(kVp)X-ray Tube Current(mA)Scan Time(s)Slice thickness or Collimation(mm)Table Speed(Table Speed(mm/rot)mm/rot)or Feed per 360 rotationor Feed per 360 rotation Pitch Pitch Interpolation Process Interpolation
19、Process Increment(mm)Increment(mm)Scan ParametersX-ray Tube VoltTable Speed&Pitch Table Speed is defined as distance traveled in mm per 360 rotationPitch =Table Feed per rotation CollimationTable Speed&Pitch Table Spee10mm 10mm P1P110mm P210mm P23030s s30s30sMore Coverage inMore Coverage inthe same
20、time with the same time with extended Pitch!extended Pitch!Pitch 2 covers Pitch 2 covers 2x2xdistance as distance as Pitch 1Pitch 110mm P110mm P230s30sMore CoverScan Range=300mm10mm P110mm P110 mm/s10 mm/s10mm P210mm P220 mm/s20 mm/s30s15sCover the same volume in shorter time with extended PitchCove
21、r the same volume in shorter time with extended PitchScan Range=300mm10mm P110mm To reduce artifacts due to table motion during spiral scanning,we use a special reconstruction process called INTERPOLATION Interpolation Interpolation Converts volume data into slice Converts volume data into slice ima