1、Chapter Potentiometric AnalysisChapter Potentiometric Analysis.1.4 Direct Potential Determination of Solution pH1.4 Direct Potential Determination of Solution pHSchematic diagram of direct potential methodReference electrodesIon meter or pH meterIndicator electrodeAgitatorWe have learned some method
2、s of measuring pH value,such as indicator liquid method,pH test paper method,the accuracy of both methods is not very high,then how to improve the accuracy of measuring pH value?Today we are going to learn a new method for measuring pH-direct potential.Principle of Determination:pH is a negative log
3、arithm of the activity of hydrogen ions,i.e.pH=-lga(H+).The pH glass electrode is used as indicator electrode(negative electrode),and the saturated calomel electrode is used as reference electrode(positive electrode).The indicator electrode and reference electrode are inserted into the cell to be te
4、sted.Working battery:Glass electrode|Liquid to be tested|glycury electrodeAt 25,the battery electromotive force is:E=(+)(-)E=SCE(K Glass 0.0592pH)=SCEK Glass+0.0592pH E=E+0.0592 pH When pH experimental conditions such as temperature are constant,the electrode potential of the pH glass electrode is l
5、inearly related to the pH of the test solution.Saturation calomel electrodeAcid-junction meterGlass film electrode.The pH scaling method It is used to determine the electromotive force of standard buffer solutions with known pH value and test fluid x with known pH value.The same reference electrode
6、and indicator electrode are used to form a working cell under the same conditions.Electromotive force of standard solutions is determined as follows:Es=Ks+2.303RTpHs/F Electromotive force of unknown solution:Ex=Kx+2.303RTpHx/FThe two equations are subtracted to obtain:Ex-Es=0.0592(pHx-pHs).Standard
7、buffer solution pH standard buffer solution must be used to determine the pH of the solution.It is the standard buffer solution.There are usually seven types of standard buffers,of which the standard buffers with pH of 6.86,9.18 and 4.00 are commonly used.(1)Application of PH standard buffer solutio
8、n:Calibration of acid meter before pH measurement.To verify the accuracy of the acid meter,for example,after calibrating the acid meter with pH6.86 and pH4.00,the pH electrode pH inserted into the solution of pH9.18 to check whether the value of the instrument is consistent with that of the standard
9、 solution.Test the performance of pH electrode.(2)The preservation of standard solutions:The standard solution should be sealed in glass or polyethylene bottles in airtight storage.Under room temperature,the standard solution should be kept for 23 months.When it is found to be cloudy,moldy or precip
10、itated,it cant be continued.The used standard solution is not allowed to be turned back.The pH value of standard buffer solution varies with temperature.Three methods of preparation of commonly used standard solutions:Potassium hydrogen phthalate standard buffer solutionc is 0.05mol/L,weigh 10.21g o
11、f potassium hydrogen phthalate(KHC8H4O4)dried at(115.05.0)for 23h,dissolve in distilled water without CO2 and dilute to 1000ml.Mixed phosphate standard buffer solutionSodium dihydrogen phosphate(Na2HPO4)and potassium dihydrogen phosphate(KH2PO4)Sodium dihydrogen phosphate Na2HPO4(3.53 0.01)g and pot
12、assium dihydrogen phosphate(KH2PO4)(3.39 0.01)g were obtained at(115.0 5.0)for 2 to 3 hours respectively.They were dissolved in distilled water which KH been boiled for 15 to 30 minutes and cooled rapidly,and diluted to 1000ml.Sodium tetraborate standard buffer solutionNa2B4O710H2O boraxC(Na2B4O710H
13、2O)is called Borax(Na2B4O710H2O)(3.80 0.01)g(note:cant be dried!)Soluble min distilled water boiled in advance for 15 to 30 minutes and cooled rapidly and diluted to 1000ml.Keep sealed in polyethylene plastic bottle.Care should be taken to prevent the entry of CO2 into the air when storing it(note:m
14、ay be used for acid meter calibration).Meter At At present,more pH gauges are used for digital display present,more pH gauges are used for digital display of the pHS-3F series acid meter.of the pHS-3F series acid meter.Method Method of use for pHS-3F acid meter:of use for pHS-3F acid meter:When meas
15、uring the pH of the solution,the pH glass When measuring the pH of the solution,the pH glass electrode and the calomel electrode are inserted into the electrode and the calomel electrode are inserted into the solution to be tested to form a battery,and the lead end of solution to be tested to form a
16、 battery,and the lead end of the electrode is connected with the pH meter.The the electrode is connected with the pH meter.The millivoltmeter is calibrated with 12 standard buffer millivoltmeter is calibrated with 12 standard buffer solution,then the test solution is measured and the pH solution,the